Gauge settings FERITSCOPE® DMP30
14.7 Default settings for decimal places
You set defaults in the gauge for the maximum number of decimal places. All newly
created batches are automatically assigned this these default variables (Preset
rounding).
To set the number of decimal places for the display values
1. > Gauge Settings > OK > More Settings > OK > Preset rounding > OK > select
desired max. number of decimal places > OK
Parameter is selected:
Parameter selection:
- Low: maximum 1 decimal place is displayed 0.9 / 9.9 / 99 / 999 / 9999
- Medium: maximum 2 decimal places are displayed 0.99 / 9.9 / 99.9 / 999 /
9999
- High: maximum 3 decimal places are displayed 0.999 / 9.99 / 99.99 / 999.9
/ 9999
2. Exit the menu: , the settings are now made.
14.8 Activating/deactivating lock code
Use the lock code to activate/deactivate the lock function (Lock Calibration) for select-
ed calibrations.
To activate the lock code
1. > Gauge Settings > OK > More Settings > Define Lock Code > OK > set desired
lock code > OK
2. Confirm the activation information: OK
3. Exit the menu: , the lock code has now been activated.
[ ]
[ ]