72
FERITSCOPE® DMP30
Calibration
9.15 Deleting all calibrations
You can delete al l your calibrations from the memory of the connected probe. This
function deletes all the calibration files and all their data!
DMP-F probe adapter: You can use this function to delete all calibrations from the
DMP-F probe adapter that were created using the connected probe. Useful when an
existing probe is to be replaced by a new one. In this case, the calibrations of the old
(e.g. defective) probe are no longer needed, so they can be deleted and do not unnec-
essarily take up memory space in the DMP-F probe adapter.
ATTENTION
Please note that a measurement without a calibration is not possible!
You must assign a new calibration to each batch already created!
Using this function you also delete the locked calibrations in the probe/DMP-
F probe adapter!
Before you start
The probe is connected to the gauge containing the memory from which you
want to delete al l calibrations.
To delete al l calibrations from the memory of the connected probe
1. Call up the delete function: > Gauge Settings > OK > Probe - Menu > OK > Delete
all calibrations > OK.
2. Start the delete process: .
3. Confirm the message: .
4. Exit the menu: , the delete process has finished.
The delete process has finished, the calibration memory of the probe is now empty.
What you should do next
Create new calibration(s), see page 63
Assign the new calibrations to the batches, page 41
[ ]