69
FERITSCOPE® DMP30
Calibration
Parameter selection:
- No: Check function deactivated
- Abs. Tol., example: ± 0.1 FN
- Rel. Tol., percentage value, example: ± 1.25 %
- Uncertainty, example: expanded measurement uncertainty for an expansion
factor k=2: ± 0.5 FN
5. Set the nominal value and the tolerance/measurement uncertainty for the cali-
bration standard you intend to use for the check (control measurement) later:
Reference Value > OK > set value > OK.
Repeat Step 5. for the tolerance/uncertainty.
When entering, pay attention to the unit of measurement. Enter the values
in the unit of measurement set for the currently open batch.
Input parameters:
- Reference Value: Nominal value of the calibration standard
- Abs. Tol./Rel. Tol. or U(k=2): Tolerance/measurement uncertainty value of
the calibration standard
6. Did you deactivate the lock function in step 2.?
- Yes: Reactivate the lock function, 69.
Function is activated:
- No: Continue with step 7.
7. Exit the menu: , the test criteria are now defined.
9.12 Lock calibration
The Lock Calibration function locks the opening and the calibration routine of the se-
lected calibration. You can unlock the menu for the locked calibration and calibration
routine only by entering the lock code. This lock function can be used only if a lock
code is defined in the gauge.
Before you start
The probe is connected to the gauge containing the calibration you want to lock.
In the gauge is the batch open to which the calibration is assigned, you want to lock.
A lock code is defined in the gauge, 90