26 Operator‘s Manual for instrument types MP0R
Data Transmission
The instrument is connected at the computer via the USB cable.
The instrument is turned ON ([OK] key).
For instrument models MP0R only: Remove the orange protective cap
from the probe.
Settings of service menu 1 must be bL = 0. Procedure 31 and 33.
Settings of service menu 4 must be SI = 0 (mean value transmission is
active). Procedure 31 and 37.
The program on the computer to which the data will transmitted is
open. Attend the manual of the program for data import.
Procedure
1. Make the desired number of measurements.
2. Set the display menu with [ ] or [ ] to show .
3. Press button [OK]. The displayed mean value of the readings is trans-
mit to the computer.
4. Delete all measurements. This step is necessary to ensure that only
the desired number of new measurements are used for the next mean
value computation. Delete procedure:
1. Set the display menu with [] or [] to show .
2. Press button [OK].
3. Set the display menu with [] to show
.
4. Press button [OK]. All measurement data will be deleted from the
measurement data memory.
RES
RES
DEL
MP0R_DataTrans_en.fm Seite 26 Donnerstag, 24. März 2011 6:46 18