Operator‘s Manual for instrument types MP0R 27
Data Transmission
Transmitting the mean values automatically to the PC while mea-
surement
In this case, all data of the measurement memory are deleted automati-
cally after the displayed mean value is transmit to the computer.
Before you start
USB driver is installed on the computer where the instrument will con-
nected.
The instrument is connected at the computer via the USB cable.
The instrument is turned ON ([OK] key).
For instrument models MP0R only: Remove the orange protective cap
from the probe.
Settings of service menu 1 must be bL = between 2 (mode SSPC: 3)
and 20 (number of readings, which are collected in one block). Proce-
dure 31 and 33.
Mode Std: Settings of service menu 4 must be SI = 0 (mean value
transmission is active, no single readings will then be transmitted).
Procedure 31 and 37.
Mode SSPC: default setting, always mean values are transmitted.
The program on the computer to which the data will transmitted is
open. Attend the manual of the program for data import.
Procedure
1. Make measurements.
2. After reaching the set number of measurements, the mean value is
sent automatically to the computer and the measurement data memory
of the instrument will be deleted. A new measurement series begin.
Please Note!
No automatically transmission of mean values in the measuring mode 90.10.