Operator‘s Manual for instrument types MP0R 47
Technical Data
PERMASCOPE MP0R and PERMASCOPE MP0R-FP - specific
technical data
ISOSCOPE MP0R - specific technical data
Applications Thickness measurement of non-magnetic coat-
ings or el. non-conductive coatings on steel or iron
(Fe).
Test method Magnetic induction
Measurement range 0 - 2500 µm (0 - 97.5 mils)
Trueness
(based on Fischer
standards)
0 ... 100 μm (0 - 3.9 mils): 1.5 μm ( 0.06 mils)
100 ... 1000 μm (3.9 - 39 mils): 1.5 % of reading
1000 ... 2500 μm (39 - 97.5 mils): 3 % of reading
Repeatability preci-
sion (based on
Fischer standards)
0 ... 100 μm (0 - 3.9 mils): 0.3 μm (0.01 mils)
100 ... 2500 μm (3.9 - 97.5 mils): 3 % of reading
Applications Thickness measurement of el. non-conductive
coatings on nonferrous metals (NFe).
Test method Eddy current
Measurement range 0 - 1200 μm (0 - 46.8 mils)
Trueness
(based on Fischer
standards)
0 ... 70 μm (0 - 2.7 mils): ± 1 μm (± 0.04 mils)
70 ... 250 μm (2 - 9.8 mils): 1.5 % of reading
250 ... 1200 μm (9.8 - 46.8 mils): 3 % of reading
Repeatability preci-
sion (based on
Fischer standards)
0 ... 50 μm (0 - 2 mils): 0.25 μm (0.01 mils)
50 ... 1000 μm (2 - 39 mils): 0,5 % of reading