48 Operator‘s Manual for instrument types MP0R
Technical Data
DUALSCOPE MP0R and DUALSCOPE MP0R-FP -
specific technical data
Applications Fe:
Non-magnetic coatings or el. non-conductive coat-
ings on steel or iron (Fe)
and
NFe:
El. non-conductive coatings on nonferrous metals
(NFe)
• Automatically recognizes the substrate material
under the coating (default setting). The DUAL-
SCOPE instruments select the appropriate test
method.
• Suited for measurements on rough surfaces.
Test method Magnetic induction and Eddy current
Measurement range Fe: 0 - 2000 μm (0 - 78 mils)
NFe: 0 - 2000 μm (0 - 78 mils)
Trueness
(based on Fischer
standards)
Fe:
0 ... 75 μm (0 - 3 mils): ≤ 1.5 μm (± 0.06 mils)
75 ... 1000 μm (3 - 39 mils): ≤ 2 % of reading
1000 ... 2000 μm (39 - 78 mils): ≤ 3 % of reading
NFe:
0 ... 50 μm (0 - 2 mils): ± 1 μm (± 0.04 mils)
50 ... 1000 μm (2 - 39 mils): ± 2 % of reading
1000 ... 2000 μm (39 - 78 mils): ≤ 3 % of reading
Repeatability preci-
sion (based on
Fischer standards)
Fe:
0 ... 50 µm (0 - 2 mils): ≤ 0.25 µm (0.01 mils)
50 ... 2000 µm (2 - 78 mils): ≤ 0.5 % of reading
NFe:
0 ... 100 μm (0 - 3.9 mils): ≤ 0.5 μm (0.02 mils)
100 ... 2000 μm (3.9 - 78 mils): ≤ 0.5 % of reading