Operator‘s Manual for instrument types MP0R 49
Technical Data
DUALSCOPE MP0RH-FP - specific technical data
Applications Fe:
Thickness measurement of non-magnetic coat-
ings or el. non-conductive coatings on steel or iron
(Fe). Measurement unaffected from the electrical
conductivity of the coating material.
and
NFe:
Thickness measurement of el. non-conductive
coatings on nonferrous metals (NFe).
Automatically recognizes the substrate material
under the coating. The DUALSCOPE instru-
ments select the appropriate test method.
Test method Magnetic induction and Eddy current
Measurement range Fe: 0 - 7000 μm (0 - 273 mils)
NFe: 0 - 2500 μm (0 - 97.5 mils)
Trueness
(based on Fischer
standards)
Fe:
0 ... 150 μm (0 - 5,9 mils): ± 5 μm (± 0.19 mils)
150 ... 3000 μm (5.9 - 117 mils): 3 % of reading
3000 ... 6000 μm (117 - 234 mils): 5 % of reading
NFe:
0 ... 50 μm (0 - 2 mils): ± 1 μm (± 0.04 mils)
50 ... 1000 μm (2 - 39 mils): ± 2 % of reading
1000 ... 2200 μm (39 - 86 mils): 3 % of reading
Repeatability preci-
sion (based on
Fischer standards)
Fe:
0 ... 200 µm (0 - 7.8 mils): 2 µm (0.08 mils)
200 ... 6000 µm (7.8 - 234 mils): 1 % of reading
NFe:
0 ... 50 μm (0 - 2 mils): 0.5 μm (0.02 mils)
50 ... 1000 μm (2 - 39 mils): 1 % of reading
1000 ... 2200 µm (39 - 86 mils): 1.5 % of reading