FLUXUS ADM 8027, F801, ADM 8127B 18 Outputs
122 UMFLUXUS_F8V4-6-1EN, 2018-10-10
18.1.1 Output Range
18.1.2 Error Value
In the following dialog, an error value can be defined which is output in case a source item cannot be measured, e.g., if
there are gas bubbles in the fluid.
Miscellaneous c-Medium sound speed of the fluid
Concentration concentration of the fluid
Signal signal amplitude of a measuring channel
SCNR ratio useful signal to correlated disturbance signal
VariAmp standard deviation of the signal amplitude
Density density of the fluid
Pressure pressure of the fluid
When configuring an analog output, the output range will be defined now. Select a list item
or other range... to enter the output range manually.
If other range... is selected, enter the values Output MIN and Output MAX. Press
ENTER after each input.
This error message will be displayed if the output range is not min. 10 % of the max. output
range. The next possible value will be displayed. Repeat the input.
example: I
MAX
- I
MIN
≥ 2 mA for a 4…20 mA current output
Tab. 18.2: Error output
error value results
Minimum the lower limit of the output range is transmitted
Hold last value the last measured value is transmitted
Maximum the upper limit of the output range is transmitted
Other value... The value has to be entered manually. It has to be within the limits of the output.
Tab. 18.1: Configuration of the outputs
source item list item output
I1 Output MAX ↕
12.0 MINIMAL