18 Outputs FLUXUS ADM 8027, F801, ADM 8127B
UMFLUXUS_F8V4-6-1EN, 2018-10-10 121
18 Outputs
If the transmitter is equipped with outputs, they have to be installed and activated before they can be used:
assign a measuring channel (source channel) to the output (if the transmitter has more than one measuring channel)
assign the physical quantity (source item) to be transmitted to the output by the source channel, and the properties of the
signal
define the behavior of the output in case no valid measured values are available
activation of the installed output in the program branch Output Options
18.1 Installation of an Output
All outputs are installed in Special Funct.\SYSTEM settings\Proc. outputs.
The source items and their scroll lists are described in Tab. 18.1.
Note! The settings will be stored at the end of the dialog. If the menu item is quit by pressing key BRK, the
changes will not be stored.
Select Special Funct.\SYSTEM settings\Proc. outputs. Press ENTER.
Select the output to be installed. Press ENTER.
The scroll list contains all actually available outputs. A tick () after a list item indicates that
this output has already been installed.
This display will be indicated if the output has not been installed yet. Select yes. Press EN-
TER.
If the output has already been installed, select no to reconfigure it or yes to uninstall the
output and to return to the previous menu item in order to select another output. Press EN-
TER.
Select in the scroll list the measuring channel to be assigned to the output as the source
channel. Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
Select a physical quantity (source item) to be transmitted to the output by the source chan-
nel.
If a binary output is configured, only the list items Limit and Pulse will be displayed.
Tab. 18.1: Configuration of the outputs
source item list item output
Measuring value physical quantity selected in the program branch Output Op-
tions
Quantity Q+ totalizer for the positive flow direction
Q- totalizer for the negative flow direction
ΣQ sum of the totalizers (positive and negative flow direction)
Limit R1 limit message (alarm output R1)
R2 limit message (alarm output R2)
R3 limit message (alarm output R3)
Pulse from abs(x) pulse without sign consideration
from x > 0 pulse for positive measured values
from x < 0 pulse for negative measured values
SYSTEM settings
Proc. outputs
Install Output
Current I1 ()
I1 enable
no >YES<
I1 disable
>NO< yes
I1 Source chan.
Channel A:
Source item
Measuring value

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F801 and is the answer not in the manual?

Flexim FLUXUS F801 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F801
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals