6 Installation of FLUXUS ADM 8027 FLUXUS ADM 8027, F801, ADM 8127B
UMFLUXUS_F8V4-6-1EN, 2018-10-10 25
6 Installation of FLUXUS ADM 8027
6.1 Location
Select the measuring point according to the recommendations in chapter 3 and 5.
Select the location of the transmitter within the range of the cable of the measuring point.
The ambient temperature at the location has to be within the operating temperature range of the transmitter and the trans-
ducers (see Technical Specification).
If the measuring point is within an explosive atmosphere, the danger zone and gases that may be present have to be de-
termined. The transducers and the transmitter have to be appropriate for these conditions.
6.2 Opening and Closing the Housing
The transmitter has a set screw which has to be unscrewed before the housing can be opened.
After the installation of the transmitter, make sure that the housings are closed correctly and that the set screw is tigh-t-
ened.
6.3 Installation
6.3.1 Wall Installation
Fix the bottom side of the upper housing to the instrument mounting plate (3) (see Fig. 6.1).
Fix the transmitter to the wall.
6.3.2 Pipe Installation
Installation on a 2" pipe
Fix the pipe mounting plate (2) to the pipe (see Fig. 6.1).
Fix the instrument mounting plate (3) to the pipe mounting plate (2) with the screws (4).
Fix the bottom side of the upper housing to the instrument mounting plate (3).
Installation on a pipe > 2"
The pipe mounting kit is fixed to the pipe with tension straps instead of the shackle (see Fig. 6.1). Push the tension straps
(5) through the holes in the instrument mounting plate (3).
6.4 Connection of the Transmitter
Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Fig. 6.1: Pipe mounting kit
Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Attention! The degree of protection of the transmitter will only be ensured if the cable glands are firmly tightened
and the housing covers are tightly screwed to the housing.
1
2
3
4
> 2 "
5
3
1
2
3
4
2 "
1 shackle
2 pipe mounting plate
3 instrument mounting plate
4 screw
5 tension strap

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F801 and is the answer not in the manual?

Flexim FLUXUS F801 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F801
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals