7 Installation of FLUXUS F801 FLUXUS ADM 8027, F801, ADM 8127B
UMFLUXUS_F8V4-6-1EN, 2018-10-10 47
7.8 Connection of the Serial Interface
The RS232 interface can only be connected outside of an explosive atmosphere because the housing has to be opened
(see Fig. 7.1).
Insert the RS232 adapter into the socket in such way that the colored lead of the cable is on the marked side of the socket.
Connect the RS232 cable to the RS232 adapter.
Connect the RS232 cable to the transmitter and to the serial interface of the PC. If the RS232 cable cannot be connected
to the PC, use the RS232/USB adapter.
The RS232 adapter, the RS232 cable and the RS232/USB adapter are part of the serial data kit (optional).
The transmitter can also be equipped with an RS485 interface (optional). For the connection see section 7.7.
For further information on the data transmission see chapter 14.
Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Note! If a problem occurs when the RS232/USB adapter is used for the connection, contact your system
administrator.
Fig. 7.1: RS232 interface of FLUXUSF801
colored lead of the RS232
adapter cable

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F801 and is the answer not in the manual?

Flexim FLUXUS F801 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F801
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals