FLUXUS ADM 8027, F801, ADM 8127B 8 Installation of FLUXUS ADM 8127B
54 UMFLUXUS_F8V4-6-1EN, 2018-10-10
Connection of the FLUXUS ADM 8127B
Connection according to IBExU07ATEX1061 (system: power supply unit - cable - transmitter)
FLUXUS ADM 8127B has an ATEX certified system approval:
The system approval allows the connection of
transmitter FLUXUS ADM 8127B
power supply FHF Bergbautechnik NG3-12ib
approved cable (max. 80 m) type L-YY(ZG)Y-2x2x0.5-60V-blue or type L-2YYC(ZG)Y-2x2x0.5
Connect the protective earth to the terminal PE of FLUXUS ADM 8127B (see Tab. 8.3).
Connect the approved cable to the terminals of the FLUXUS ADM 8127B and the terminals of the power supply unit.
For a drawing of the system see annex E.
8.7 Connection of the Outputs
Select the cable gland for mines for the connection of the outputs. After the installation, the cable has to fit firmly in the
cable gland for mines:
M25 (9/12) for cable diameters 9...12 mm
M25 (14/16) for cable diameters 14...16 mm
Prepare the output cable with the cable gland for mines.
Remove as little of the insulation as possible when stripping the cable. The insulation has to cover the lead up to the ter-
minal.
Push the output cable through the screwed socket (1), the compression rings (2), the sealing ring (3) and the gland body
(4) (see Fig. 8.7).
Insert the output cable into the housing.
Screw the gland body (4) into the housing (see Fig. 8.7).
Fix the screwed socket (1) and the gland body (4) to the clamp by firmly tightening the screws of the clamp.
Connect the leads to the terminals of the transmitter (see Fig. 8.3, Fig. 8.8 and Tab. 8.3).
Tab. 8.2: Connection of the intrinsically safe power supply
terminal connection
PE earth
L+ 12 V DC (+), U
i
= 13.2 V
L- 12 V DC (-), U
i
= 13.2 V
IM2 SYST Ex ib I
-20 °C Ta 50 °C
IBExU07ATEX1061
Tab. 8.3: Connection according to IBExU07ATEX1061
terminal connection
PE earth
L+ terminal (+) on the power supply
L- terminal (-) on the power supply
Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F801 and is the answer not in the manual?

Flexim FLUXUS F801 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F801
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals