FLIR DM286 USER MANUAL Document Identifier: DM286-en-US_AA
14. Specifications
14.1 General specifications
Maximum voltage 1000 V DC or 1000 V AC RMS
Display Counts 6000
Polarity Indication Automatic, positive implied, negative indicated
Over-range Indication OL
Measuring Rate 3 samples per second
Bluetooth range 32 ft. (10 m)
Power Requirements TA04 rechargeable lithium-polymer battery (3.7 V, 3050 mAh)
Battery life Approx. 6.5 hours in IGM mode with display brightness set to
medium
Approx. 10 hours in Multimeter mode (medium brightness)
Auto Power Off For Meter: set to OFF, 2, 5, or 10 minutes
For Worklight only: set to OFF, 5, 15, or 30 minutes
Operating Temp/RH 14°F to 86°F (-10°C to 30°C), < 85% RH
86°F to 104°F (30°C to 40°C), < 75% RH
104°F to 122°F (40°C to 50°C), <45% RH
Storage Temperature/RH -4°F to 140°F (-20°C to 60°C), 0 to 80% RH (without batteries)
Temperature Coefficient 0.1 x (specified accuracy)/°C, < 64.4°F (18°C), > 82.4°F (28°C)
Operating Altitude 6560 ft. (2000 m)
Calibration Cycle One year
Weight 1.25 lbs. (569 g) without battery (approx.)
Dimensions 7.8 x 3.6 x 2.0 in. (19.9 x 9.2 x 5.1 cm)
Safety Complies with IEC 61010-1 CAT IV-600 V, CAT III-1000V
Certifications: ETL, FCC, CE, RCM, UKCA, FDA, EN60825,
EN50689
EMC: EN 61326-1; pollution degree: 2
IP40 rated
Drop protection: 6.6 ft. (2m)