FLIR DM286 USER MANUAL Document Identifier: DM286-en-US_AA
14.2 Thermal MSX Imaging Specifications
IR temperature measurement range 23 to 572°F (-5 to 300°C)
OL is display when range is exceeded
IR measurement stabilization
The approximation symbol (~) is shown next to the
temperature reading when the reading accuracy has
not yet stabilized (~125
o
C, for example). The symbol will
disappear when the reading is stable. Stabilization
typically requires 2 to 3 minutes (at room temperature).
IR temperature resolution 0.1°
Image sensitivity ≤ 150mK (0.15°C)
IR temperature accuracy
± 5.4°F (3°C) or 3% (whichever is greater) for
temperatures > 77°F (25°C).
± 9°F (5°C) for temperatures 14 to 77°F (-10 to 25°C).
IR image capture frequency 9 Hz
Emissivity 0.10 to 1.00 (default is 0.95); select from four presets or
set manually
Response time 150ms
Spectral Response 8 to 14um
Scanning and targeting Continuous scan. Crosshairs and laser pointer targeting.
Repeatability 0.5%
Image Detector FLIR Lepton® microbolometer
MSX Multi-Spectral Imaging (digital camera adds visible light
details to the thermal image for clarity)
Field of View (FOV) 160 x 120 pixels (57
o
x 44
o
)
Colour Palettes Iron, rainbow, arctic, white hot, and black hot
Stored thermal image radiometry Fully radiometric
Laser type Class 1 red
Laser power < 0.39mW
Laser wavelength 640 to 660 nm
14.3 Digital Camera Specifications
Digital camera resolution 2 MP (1600 x 1200 pixels)
Focus Fixed
Field of view 71° x 56°