Correction (NUC) Types
One Point
Sets the gain terms to “1” and computes the offset terms.
Uses a single NUC source. Does not compute a BP
correction.
Two Point
Sets both the gain and offset terms. Uses two NUC
sources. Computes a bad pixel correction.
Offset Update
Retains the current NUC gain terms and updates the offset
terms. Uses a single NUC source. Retains the current bad
pixel (BP) correction.
Correction Sources
Internal Flag
Use the internal flag as the NUC source. This is disabled
for 2-point NUC because the flag is not temperature
controlled
External Blackbody
Use an external blackbody as the NUC source. Program
will prompt the user to place each source in front of the
camera. NUC source needs to fill the entire field of view.