43B
Service Manual
5-16
To exit and save the INVALID calibration data:
Press
YES.
The test tool will show the message
The test tool needs calibration. Please contact
your service center at power on. The calibration date and number will not be
updated. A complete recalibration must be done.
To exit and maintain the old calibration data:
Turn the test tool off.
No power adapter voltage
WARNING.No adapter present.
Calibration data will not be saved.
Exit maintenance mode?
To save the calibration data:
Press
NO
The test tool returns to the maintenance mode. Connect a correct power adapter,
and press
to exit and save.
To exit without saving the calibration data:
Press
YES

Table of Contents

Other manuals for Fluke 43B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 43B and is the answer not in the manual?

Fluke 43B Specifications

General IconGeneral
CategoryPower Quality Analyzer
InterfaceRS-232
Volts AC/DCYes
OhmsYes
ContinuityYes
Diode TestYes
CapacitanceYes
Harmonics MeasurementUp to 51st harmonic
DisplayLCD
Battery LifeApproximately 6 hours
Weight1.1 kg (2.4 lb)
Data StorageYes
Amps AC/DC0.1 to 1000 A (with clamp)
Power MeasurementW, VA, VAR, PF
Dimensions50 mm
Current Measurement RangeVia current clamp accessories (not built-in)

Related product manuals