43B
Service Manual
5-16
To exit and save the INVALID calibration data:
Press
YES.
The test tool will show the message
The test tool needs calibration. Please contact
your service center at power on. The calibration date and number will not be
updated. A complete recalibration must be done.
To exit and maintain the old calibration data:
Turn the test tool off.
No power adapter voltage
WARNING.No adapter present.
Calibration data will not be saved.
Exit maintenance mode?
To save the calibration data:
Press
NO
The test tool returns to the maintenance mode. Connect a correct power adapter,
and press
to exit and save.
To exit without saving the calibration data:
Press
YES