EasyManuals Logo

Fluke 43B Service Manual

Fluke 43B
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #94 background imageLoading...
Page #94 background image
43B
Service Manual
5-10
ST8004.WMF
Figure 5-4. 5500A Scope Output to Input 1
3. Set the 5500A to source a 1V, 1 MHz fast rising (rise time 1 ns) square wave
(SCOPE output, MODE edge).
4. Set the 5500A to operate (OPR).
5. Press
to start the calibration.
The Delta T gain, Trigger Delay (CL0720), and Pulse Adjust Input 1 (CL0640) will
be calibrated.
6. Wait until the display shows
Pulse Adj A (CL 0640):READY.
7. When you are finished, set the 5500A to Standby.
8. Continue at Section 5.6.3.
5.6.3 Gain DMM (Gain Volt)
Warning
Dangerous voltages will be present on the calibration source
and connection cables during the following steps. Ensure that
the calibrator is in standby mode before making any connection
between the calibrator and the test tool.
Proceed as follows to do the Gain DMM calibration.
1. Press
to select the first calibration step in Table 5-3.
2. Connect the test tool to the 5500A as shown in Figure 5-5.

Table of Contents

Other manuals for Fluke 43B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 43B and is the answer not in the manual?

Fluke 43B Specifications

General IconGeneral
CategoryPower Quality Analyzer
InterfaceRS-232
Volts AC/DCYes
OhmsYes
ContinuityYes
Diode TestYes
CapacitanceYes
Harmonics MeasurementUp to 51st harmonic
DisplayLCD
Battery LifeApproximately 6 hours
Weight1.1 kg (2.4 lb)
Data StorageYes
Amps AC/DC0.1 to 1000 A (with clamp)
Power MeasurementW, VA, VAR, PF
Dimensions50 mm
Current Measurement RangeVia current clamp accessories (not built-in)

Related product manuals