Performance Verification
4.8 Sags & Swells
4
4-29
8. Press to confirm; mark changes to .
9. Press
to highlight START.
10. Press
to start the measurement.
11. Set the 5500A to 1.5V DC.
Now the measurement should start and continue for 10s. The trigger point is after 2
divisions. With
you can move a cursor and with you can toggle between the
cursors.
Checking the result:
1. Press
to move the left cursor to the fourth division.
2. Press
to toggle to the right cursor.
3. Press to move the right cursor to the fifth division.
4. Observe the readings and verify that they are between 1.38 and 1.62 kA.
5. When you are finished, set the 5500A to Standby
4.8 Sags & Swells
Proceed as follows to test the sags & swells mode.
ST8001.WMF
Figure 4-9. Test Tool Input 1 & 2 5500A Normal Output
1. Connect the test tool to the 5500A as shown in Figure 4.9.
2. Press
.
3. Press
to highlight SAGS & SWELLS.
4. Press
to enter mode.
Now the RECORD TIME is highlighted. If the time is not 4 minutes then:
1. Press
2. Press to highlight 4 minutes.
3. Press
to confirm; mark changes to .