43B
Service Manual
4-32
4.12 Transients.
Proceed as follows to test the TRANSIENTS function:
1. Connect the test tool to the 5500A as show in Figure 4-12.
ST8586.WMF
Figure 4-12. Test Tool Input 1 to 5500A Normal Output
2. Press .
3. Press
to highlight TRANSIENTS.
4. Press
to select the TRANSIENTS mode.
VOLTAGE CHANGE will be highlighted, continue with:
1. Press
.
2. Press
to select 20%.
3. Press
to confirm; mark changes to ■.
4. Press
to select START.
5. Set the 5500A to 20V, 60Hz (NORMAL output, WAVE sine).
6. Press
to start the test. No transients should be captured.
7. Set the 5500A to 22.5V.
8. Now transients should be captured. After 40 transients (maximum) the 40th transient
will be visible.
9. When you are finished, set the 5500A to Standby.