EasyManua.ls Logo

Fluke 43B - Input 2 Frequency Measurement Accuracy Test

Fluke 43B
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verification
4.5 Input 1 and Input 2 Tests in the SCOPE MODE.
4
4-9
4.5.4 Input 2 Frequency Measurement Accuracy Test
Proceed as follows to test the Input 2 frequency measurement accuracy:
1. Connect the test tool to the 5500A as shown in Figure 4-3.
ST8588.wmf
Figure 4-3. Test Tool Input 2 to 5500A NORMAL output
2. Select the AUTO test tool setup:
Press
to select the MENU.
Press
till SCOPE is highlighted.
Press
to select SCOPE mode
3. Select the following test tool setup:
Press to select menu SCOPE SETUP.
Press
to highlight Input 2 Reading
Press
to go to Input 2 READING.
Press
to highlight Hz.
Press
to confirm; mark changes to .
Press
to highlight Input 2 Coupling.
Press
to select the Input 2 Coupling menu.
Press
to highlight DC Coupling.
Press
to confirm; mark changes to
Press
to return to SCOPE.
4. Set the 5500A to source a sine wave of 600 mV, 15 kHz (NORMAL output, MODE
WAVE sine).
5. Observe the Input 2 main reading on the test tool and check the reading between 14.8
and 15.2 kHz.
6. When you are finished, set the 5500A to Standby.

Table of Contents

Other manuals for Fluke 43B

Related product manuals