Performance Verification
4.5 Input 1 and Input 2 Tests in the SCOPE MODE.
4
4-13
8. Select timebase of 10 µs/d.
Press
to select RANGE.
Press
to select 10 µs/d.
9. Set the 5500A to source a 20 kHz leveled sine wave of 100 mV peak-to-peak (if
necessary adjust the 5500A to half a division peak-to-peak on the display).
10. Verify that the signal is well triggered , if necessary adjust the trigger level (see 7).
11. When you are finished, set the 5500A to Standby.
4.5.7 Input 1 Trigger Level and Trigger Slope Test
Proceed as follows:
1. Connect the test tool to the 5500A as shown in Figure 4-4.
ST8586.WMF
Figure 4-4. Test Tool Input 1 to 5500A Normal Output
2. Select the AUTO test tool setup:
Press
to select the MENU.
Press
till SCOPE is highlighted.
Press
to select SCOPE mode
3. Select the following test tool setup:
Press to select menu SCOPE SETUP.
Press
to select Input 1 READING.
Press
to highlight AC+DCrms.
Press
to confirm; mark changes to .
Press
to highlight Input 1 Coupling.
Press
to select the Input 1 Coupling menu.
Press
to highlight DC Coupling.
Press
to confirm; mark changes to .
Press
to return to SCOPE.