EasyManuals Logo

Fluke 43B Service Manual

Fluke 43B
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #97 background imageLoading...
Page #97 background image
Calibration Adjustment
5.6 Final Calibration
5
5-13
ST8003.WMF
Figure 5-7. Four-wire Ohms calibration connections
3. Set the 5500A to the first test point in Table 5-4. Use the 5500A “COMP 2 wire”
mode for the calibration adjustments up to and including 100 k. For the higher
values, the 5500A will turn off the “COMP 2 wire” mode.
4. Set the 5500A to operate (OPR).
5. Press
to start the calibration.
6. Wait until the display shows the calibration status
:READY.
7. Press
to select the next calibration step, set the 5500A to the next calibration
point, and start the calibration. Continue through all calibration points.
8. When you are finished, set the 5500A to Standby.
9. Continue at Section 5.6.6.
Table 5-4. Ohm Gain Calibration Points
Cal Step Input Value
Gain Ohm (CL 0860) [Cap. Pos. (CL 0920), Cap.Neg. (CL 0921)]
1)
100
Gain Ohm (CL 0861) [Cap. Pos. (CL 0922), Cap.Neg. (CL 0923)]
1)
1 k
Gain Ohm (CL 0862) [Cap. Pos. (CL 0924), Cap.Neg. (CL 0925)]
1)
10 k
Gain Ohm (CL 0863) [Cap. Pos. (CL 0926), Cap.Neg. (CL 0927)]
1)
100 k
Gain Ohm (CL 0864) 1 M
Gain Ohm (CL 0865) [Gain Ohm (CL 0866)]
2)
10 M
1)
The capacitance measurement current calibrations (Cap.Pos. and Cap.Neg) are done automatically after
the Gain Ohm calibration.
2)
The Gain Ohm (CL0866) calibration step is done automatically after the Gain Ohm (CL0865) calibration.
5.6.6 Capacitance Gain Low and High
Proceed as follows to do the Capacitance Gain calibration:
1. Press
to select calibration adjustment step Cap. Low (CL 0900):IDLE
2. Connect the test tool to the 5500A as shown in Figure 5-8.

Table of Contents

Other manuals for Fluke 43B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 43B and is the answer not in the manual?

Fluke 43B Specifications

General IconGeneral
CategoryPower Quality Analyzer
InterfaceRS-232
Volts AC/DCYes
OhmsYes
ContinuityYes
Diode TestYes
CapacitanceYes
Harmonics MeasurementUp to 51st harmonic
DisplayLCD
Battery LifeApproximately 6 hours
Weight1.1 kg (2.4 lb)
Data StorageYes
Amps AC/DC0.1 to 1000 A (with clamp)
Power MeasurementW, VA, VAR, PF
Dimensions50 mm
Current Measurement RangeVia current clamp accessories (not built-in)

Related product manuals