Specifications
95
Specifications
Environmental Specifications
Operating temperature and relative
humidity
32ºF to 113ºF (0ºC to +45ºC)
Note
The battery will not charge if the internal temperature of the tester is
above 113
º
F (45ºC).
Operating relative humidity
(% RH without condensation)
90% (50ºF to 95ºF; 10ºC to 35ºC)
75% (95ºF to 113ºF; 35ºC to 45ºC)
Storage temperature -4ºF to 140ºF (-20ºC to +60ºC)
Shock and vibration Random, 2 g, 5 Hz-500 Hz (Class 2) 1 m drop test
Safety IEC 61010-1: CAT none, Pollution degree 2
Altitude 4,000 m; Storage: 12,000 m
EMC FCC Part 15 Class A, IEC 61326-1: Portable, South Korean Standards
South Korean EMC Standards
Conforms to relevant South Korean EMC Standards.
Electromagnetic Compatibility.
Applies to use in Korea only. Class A
Equipment (Industrial Broadcasting &
Communications Equipment) [1]
[1] This product meets requirements for industrial (Class A) electromagnetic wave
equipment and the seller or user should take notice of it. This equipment is
intended for use in business environments and is not to be used in homes.