EasyManua.ls Logo

Fourtec DaqPRO 5300 - Page 12

Fourtec DaqPRO 5300
99 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Chapter 1 DaqPRO
Connect the sensor to the terminal block at the top of DaqPRO:
Figure 2: DaqPRO’s inputs block terminal
Sensors must be added successively, starting with input1. If a single sensor is used it must be
connected to Input1. If two sensors are used, they must be connected to Input-1 and Input-2 and so on.
Alarm Output
I/O8 (Input/Output8) serves either as an input or as alarm output.
Polarity
Current, voltage, thermocouples and user defined sensors have distinct polarity. Be careful to connect
them in the right polarity.
Frequency/Pulse Counter
Connect the signal wires to I/O8 screw terminals, and select Frequency or Pulse counter for Input 1
from the Setup menu. Inputs 2 to 7 are still available for other sensors.
The Frequency/pulse counter is optically isolated from the internal circuitry and can simultaneously
measure a signal source, together with another input.
3-wire PT 100
You have to use two inputs to connect a 3 wire PT 100. You can connect one 3-wire PT 100 to input1
and input5, and/or inputs 2 and 6, and/or inputs 3 and 7 and/or inputs 4 and 8.
Connect the single wire to the plus (+) terminal of the first input and the common end wires to the minus
(-) terminals of both the inputs.
www.esis.com.au
Ph 02 9481 7420
Fax 02 9481 7267
esis.enq@esis.com.au

Table of Contents