84 Chapter 4 Specifications
Chapter 4 Specifications
DaqPRO
Inputs
8 channel differential inputs
Selectable type for each
input: 0-24 mA, 0-50 mV,
0-10 V, NTC, PT-100,
Thermocouple,
Pulse counter (Input 1 only),
Frequency (Input 1 only) and
User Defined sensors
Input types
0 24 mA
Range 0 to 24 mA
Resolution 0.47µA
Accuracy ±0.5 %
Loop impedance 21
Maximum load 30 mA, 5.2 V
0 50 mV
Range 0 to 50 mV
Resolution 3 µV
Accuracy ±0.5 %
Input impedance 50 M
Maximum voltage 5.2 V
0 10 V
Range 0 to 10 V
Resolution 200 µV
Accuracy ±0.5 %
Input impedance 125 K
Maximum voltage 25 V
Temperature NTC-1
NTC 10 K resistor
Range -25 to 150 ºC
Resolution 0.05 ºC
Accuracy ±0.5 %
Temperature NTC-2
NTC 100 K resistor
Range -25 to 150 ºC
Resolution 0.05 ºC
Accuracy ±0.5 %
Temperature PT100
Range -200 ºC to 400 ºC
Resolution 0.1ºC (7 m)
Accuracy -200 to -50 ºC ±0.5 %
50 to 400 ºC ±0.5 %
-50 to 50
ºC ±0.5 ºC
Possible to connect up to 8
PT100 2-wires or 4 PT100 3-
wires
Temperature Thermocouple J
Range -200 ºC to 1,200 ºC
Resolution 0.1 ºC (1 µV)
Accuracy -200 to -50 ºC ±0.5 %
50 to 1200 ºC ±0.5 %
-50 to 50 ºC ±0.5 ºC
Cold junction compensation ±0.3 ºC
Temperature Thermocouple K
Range -250 to 1200 ºC
Resolution 0.1 ºC (1 µV)
Accuracy -250 to -50 ºC ±0.5 %
50 to 1200 ºC ±0.5 %
-50 to 50 ºC ±0.5 ºC
Cold junction compensation ±0.3 ºC
Temperature Thermocouple T
Range -200 to 400 ºC
Resolution 0.1 ºC (1 µV)
Accuracy -200 to -50 ºC ±0.5 %
50 to 400 ºC ±0.5 %
-50 to 50 ºC ±0.5 ºC
Cold junction compensation ±0.3
ºC
www.esis.com.au
Ph 02 9481 7420
Fax 02 9481 7267
esis.enq@esis.com.au