Deactivate "Encrypt password (Twincat 3 only)"
If connection is successful, an X is displayed in the overview.
5
Close the window by clicking "Close"
Possible errors/points to check if the route cannot be set as described above in-
clude:
a)
Check network settings:
The calibration station must be accessible in the Windows command prompt
(CMD) at the IP address 192.168.120.50 with the ping command.
b)
Various security programmes (e.g. F-Secure, etc.) may block the network
connection.
If necessary, allow network traffic
c)
Reset the calibration station and ensure that the green LED is on
d)
Remove the USB cable for a short time, then reconnect it and try again
e)
Restart the PC