Calibrating the
TransPocket
welding system
IMPORTANT! Before starting the calibration process:
-
The power source to be calibrated must be connected to the calibration sys-
tem.
-
The power source to be calibrated must be switched on at least 5 minutes
before the calibration process starts.
The calibration process is started from the calibration software on the com-
puter / laptop.
1
Set manual metal arc welding mode on the power source
2
Start the calibration program
3
Click [New calibration]
The preliminary settings / contractor data are displayed.
4
Enter the details for a new contractor (= Person performing calibration):
-
Name
-
Address
-
Postcode
-
Calibration technician
5
Click on [Add]
or
select current contractor from the list shown
6
Enter the serial numbers of the measuring equipment:
-
Calibration system
-
Welding wire calibration system
-
Gas measurement system
7
Click [Next >]
The preliminary settings / customer data are displayed.
8
Enter the details for a new client:
-
Name
-
Address
-
Postcode
-
Customer number
9
Click on [Add]
or
select current client from the list shown
Under Others, inventory numbers, order numbers and observations can be
entered.
67
EN

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fronius Calibration system 2.0 and is the answer not in the manual?

Fronius Calibration system 2.0 Specifications

General IconGeneral
BrandFronius
ModelCalibration system 2.0
CategoryTest Equipment
LanguageEnglish