EasyManuals Logo

Fronius Calibration system 2.0 User Manual

Fronius Calibration system 2.0
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
10
Click [Next >]
The preliminary settings / device selection are displayed.
11
Under Settings, select:
o Fronius devices
12
Select the respective TransPocket variant
13
Enter the serial number and name of the power source
14
Click [Next >]
The preliminary settings / calibration parameters are displayed:
five measured values M1 - M5 for each power source
-
Welding current [A]
15
Click [Next >]
The preliminary settings / general measurement data are displayed.
16
Click [Start measuring process]
The following data is displayed:
-
Ambient temperature [° C]
-
Mains voltage [V]
-
Open circuit voltage U
S
[V]
IMPORTANT! The mains voltage displayed is incorrect!
With single-phase machines, the mains voltage must be measured manually
using a FLUKE measuring device, and entered!
17
Measure the mains voltage
18
Enter the mains voltage measured
19
[ ] Enable In Use and enter the measuring device used under External meas-
uring equipment
20
Carry out a visual inspection of the welding system and the calibration sys-
tem, according to the calibration instructions
21
If the visual inspection does not reveal any defects, select [Click here]
22
Click [Next >]
The summary of the calibration process is displayed.
23
Define the calibration class:
o Precision
o Standard
24
Set the specified measured value M1 on the power source
68

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fronius Calibration system 2.0 and is the answer not in the manual?

Fronius Calibration system 2.0 Specifications

General IconGeneral
BrandFronius
ModelCalibration system 2.0
CategoryTest Equipment
LanguageEnglish