No USB/Speed-
Net connection
No USB/SpeedNet connection between power source and calibration software
USB connection error
Check that the USB Ethernet adapters are properly plugged into the
power source and calibration system.
Check the Ethernet connection cable for any possible damage and
for functionality.
Check that the Ethernet connection cable is connected correctly to
the USB Ethernet adapters.
Check that the DHCP server is running (icon in the task bar!). If the
DHCP server is not running, restart the software and check the net-
work settings.
Disconnect the USB briefly from the power source and reconnect it
again.
Cause:
Remedy:
94