Calibrating the
welding system
IMPORTANT! Before starting the calibration process:
-
All devices to be calibrated must be connected to the calibration system:
power source, wirefeeder, shielding gas supply; etc.
-
The power source to be calibrated must be switched on at least 5 minutes
before the calibration process starts.
The calibration process is started from the calibration software on the com-
puter / laptop.
1
Start the calibration program
2
Click [New calibration]
The preliminary settings / contractor data are displayed.
3
Enter the details for a new contractor (= Person performing calibration):
-
Name
-
Address
-
Postcode
-
Calibration technician
4
Click on [Add]
or
select current contractor from the list shown
5
Enter the serial numbers of the measuring equipment:
-
Calibration system
-
Welding wire calibration system
-
Gas measurement system
6
Click [Next >]
The preliminary settings / customer data are displayed.
7
Enter the details for a new client:
-
Name
-
Address
-
Postcode
-
Customer number
8
Click on [Add]
or
select current client from the list shown
Under Others, inventory numbers, order numbers and observations can be
entered.
9
Click [Next >]
The preliminary settings / device selection are displayed.