EasyManua.ls Logo

Fronius Calibration system 2.0 - Calibrating Transtig 1750 Puls

Fronius Calibration system 2.0
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrating Tran-
sTig 1750 Puls
IMPORTANT! Before starting the calibration process:
-
All devices to be calibrated must be connected to the calibration system:
power source, shielding gas supply, etc.
-
The power source to be calibrated must be switched on at least 5 minutes
before the calibration process starts.
1
Connect the Tuchel adapter cable to the power source and to the calibration
system
This enables the calibration process to be started from the calibration software
on the computer/laptop.
2
Set manual metal arc welding mode on the power source and in the Rod elec-
trode setup menu, specify the welding parameter Hti = 0
1
2
3
Start the calibration program
4
Click [New calibration]
The preliminary settings / contractor data are displayed.
5
Enter the details for a new contractor (= Person performing calibration):
-
Name
-
Address
-
Postcode
-
Calibration technician
6
Click on [Add]
or
select current contractor from the list shown
7
Enter the serial numbers of the measuring equipment:
-
Calibration system
-
Gas measurement system
70

Table of Contents