Chapter 5: Diagnostics, maintenance, and testing 
 
186  Vigilant VS1 and VS2 Technical Reference Manual 
Starting and stopping a walk test 
A walk test lets you test individual zones or devices without having to create an 
actual alarm condition. You can conduct a walk test in silent or audible mode. In 
silent mode, the audible devices (NACs) do not sound. 
Zones should be placed in walk test one at a time. This allows the balance of the 
system to remain in service. When a device is activated for test, it remains active 
throughout the test and does not restore until walk test is ended. When walk test is 
ended, all of the tested devices restore. 
Note: Walk test will not operate with an active alarm or supervisory event at the 
panel. 
When you enter walk test mode: 
•  The Test LED flashes for an audible walk test and is steady for a silent walk test. 
•  Walk Test is displayed on the LCD. Walk Test is replaced with the device activation 
message during the period the device is activated while in walk test. 
•  The panel enters a trouble state. There is no fire protection for the zone in walk 
test. If an unselected zone or device goes into alarm or trouble, all programmed 
outputs operate as programmed. 
Zones and devices behave as follows during walk test. 
•  For alarm events, the appropriate panel and annunciator LEDs and buzzers turn 
on. In the audible (NAC) test mode a four-second pulse sounds for NAC types: 
continuous nonsilenceable, continuous silenceable, Genesis nonsilenceable, 
Genesis with audible silence only, and Genesis with audible and visible silenceable 
on the audible devices. A single 3-3-3 cycle sounds for NAC types: temporal 
nonsilenceable, temporal silenceable, coder basic, coder continuous, coder 
120spm, and coder temporal on the audible devices. 
•  For trouble events, the appropriate LEDs and the buzzers turn on. In the audible 
(NAC) test mode a two-second pulse sounds on the audible devices. 
•  For ground fault events, the appropriate LEDs and the buzzers turn on. In the 
audible (NAC) test mode, a two-second pulse sounds on the audible devices. 
•  All device activations are latched until walk test is ended or the panel is reset. 
When walk test is ended, it takes approximately eight seconds for the devices to 
restore. If the device being tested is in a zone and does not restore, the zone does 
not reset. 
•  If auxiliary power is resettable, the auxiliary power is interrupted while the zone is 
reset.