EasyManua.ls Logo

GE DR60 - Page 115

GE DR60
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DR60
DR60-TM-EN-2A
115
AC and DC voltage interruptions
Test level: 0% residual voltage
Duration time
a.c.: 300 cycles
d.c.: 5s
Conducted RF
immunity, 0 to
150 kHz
IEC 61000-4-16:
2009
Zone A
Test voltage
Differential mode: 150 V rms
Common mode: 300 V rms
Voltage ripple
IEC 61000-4-
17:1999
Test level: 15 % of rated d.c. value
Test frequency: 120Hz, sinusoidal
waveform.
Damped
oscillatory wave
immunity test
IEC 61000-4-
18:2006
Voltage oscillation frequency: 1MHz
Differential mode: 1kV peak voltage;
Common mode 2,5kV peak voltage
---
Gradual Startup
Shut-down ramp: 60s
Power off: 5m
Start-up ramp: 60s
Radio-frequency
disturbance
CISPR11:2009
Radiated emission below 1GHz
class A
30 MHz to 230 MHz
40 dB(µV/m) quasi peak at 10 m
230 MHz to 1 000 MHz
47 dB(µV/m) quasi peak at 10 m
Radio
disturbance
CISPR22:2008
Radiated emission above 1GHz
1 GHz to 3 GHz
56 dB(µV/m) Average
76 dB (µV/m) peak at 3m
3 GHz to 6 GHz
60 dB(µV/m) Average
80 dB (µV/m) peak at 3m

Table of Contents