Printing and recording
8-5
Recording on alarms
An automatic strip chart recording is activated when the following alarms reach the red level:
Asystole, Tachy, Brady, Art High, Art Low, V Fib, and V Tachy. When recording is activated by
alarms, the recording time is always 30 seconds and the delay always 12 seconds.
1. Press the
Print/Record key.
2. Select Record Waveforms.
3. Select Start on Alarms and Yes.
The following alarms start recording:
All alarm recordings are marked with the alarm source.
Alarm Recorded parameters
Asystole
Tachy/Brady
Art High/Low
V Fib
V Tachy
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
Arrhythmia analysis Alarms which start the recording (if the alarm
is set on/level red)
Severe Asystole, Tachy/Brady, Art High/Low, V Fib, V Tachy
Print/
Record