B30 Patient Monitor
16-6
Autocycling
The Auto On/Off key sets automatic NIBP measurement at selected intervals on and off.
Autocycling is synchronized to real time so that if the first measurement was at 12.02, the next
measurement is at 12.05 and again at 12.10 (5 min. interval.)
To start the autocycling, do one of the following:
Press the Auto On/Off module or side panel key, or
press the
NIBP key on the Command Board and select Start Cycling.
The bar at the bottom of the NIBP field shows the time remaining to the next
measurement.
To stop autocycling:
Press the Auto On/Off module or side panel key or
press the
NIBP key on the Command Board and select Stop Cycling.
Setting cycle time
The possible intervals for autocycling are 1, 2.5, 3, 5, 10, 15, 30 minutes or 1, 2 or 4 hours.
To set the cycle time:
Press the NIBP key on the Command Board and select Cycle Time.
Select the alternative with the ComWheel.
NIBP