B30 Patient Monitor
10-8
Table 10-8 Other problems related to NIBP measurement
Table 10-9 Other problems related to temperature measurement
Table 10-10 Other problems related to SpO
2
measurement
Other situations
The following table lists some other situations that may occur during monitoring and possible
explanations.
Table 10-11 Other operation problems
Symptom Possible cause and solution
NIBP
measurement
does not work or
values seem
unstable
Check that cuff tubings are not bent, stretched, compressed or loose.
Prevent motion artifacts.
Use cuffs of correct size.
Symptom Possible cause and solution
Temperature
measurement
fails
Check that you are using a correct probe.
Try another probe.
Symptom Possible cause and solution
SpO
2
signal is
poor
Check the sensor and sensor positioning.
Change the averaging time from slow to normal.
Note that skin pigment causes differences.
Make sure that the patient is not shivering.
Symptom Possible cause and solution
Printing is not
possible
Printer selection is None; change it through Print/Record - Printer
connection.
Printer is not connected to the network. Check printer cable.
The measured
values are not
displayed
Check that you have selected the desired parameter to a waveform
or digit field.
The monitor does
not start
Check that the power cord is properly connected.
Check that all cables are properly connected.
Check the fuses and replace them if necessary.
You cannot
perform a
measurement or
a function
Check that the measurement module is properly installed.
Check that all cables are properly connected.
Remove the module and reinstall it.