EasyManua.ls Logo

GE Healthcare B30 - Page 153

GE Healthcare B30
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
10-7
Table 10-4 Other problems related to batteries
Table 10-5 Other problems related to ECG measurement
Table 10-6 Other problems related to impedance respiration measurement
Table 10-7 Other problems related to InvBP measurement
Symptom Possible cause and solution
Battery operation
time is markedly
shortened
Condition the batteries according to the instructions in this
manual.
Symptom Possible cause and solution
ECG signal is
noisy or no QRS is
detected
Ensure that the patient is not shivering.
Incorrect ECG filter.
Check the filter through ECG - ECG Setup - Filter.
Poor electrode quality or wrong positioning.
Check the electrodes and cables and their placement. See
section“ECG” for details.
Change the lead.
Remove the ECG cable from the connector and reinsert it.
Pacer markers are
not visible
Check that:
- The pacer markers have been selected ON.
- The pacemaker has been adjusted correctly and not above R.
- The pacemaker functions correctly: ECG cables, electrodes and
setups are correct.
Thick ECG
baseline
ECG cable is looped.
Other electrical power cables are near the ECG leadwires.
Incorrect ECG filter.
Incorrect power frequency of the monitor.
Symptom Possible cause and solution
Respiration
measurement
fails
Check the electrode quality and positioning.
Adjust the detection limits. During ventilator-supported breathing, the
respiration calculation may count only ventilator-produced
inspirations and expirations.
Other electrical devices may interfere with the measurement.
Symptom Possible cause and solution
InvBP readings
seem unstable
Make sure there are no air bubbles in the transducer system. Flush
and zero.
Place the transducer on the patient’s mid-heart level and zero.

Table of Contents

Related product manuals