B30 Patient Monitor
4-16
Reminder volume
1. Press the Monitor Setup key.
2. Select Install/Service and enter the password.
3. Select Installation - Alarm Options.
4. Select Reminder Volume and adjust the alarm reminder volume with the ComWheel.
Changing the tone pattern
The monitor has three choices of alarming tone patterns: ISO, ISO2 and General.
To change the tone pattern:
1. Press the
Monitor Setup key.
2. Select Install/Service and enter the password.
3. Select Installation - Alarm Options.
4. Select Alarm Tones and ISO, ISO2 or General.
Monitor
Setup
Monitor
Setup