B30 Patient Monitor
4-12
Recording time is 30 seconds consisting of 12 seconds recording from the recorder memory
and 18 seconds real-time recording. The alarm source is always marked to the alarm
recordings. The following alarms start the recording:
Showing alarm history
Press the Pt. Data & Trends key and select Alarm History. This displays a list of the
last 20 alarms that have reached the yellow or the red alarm level. The time and type of
occurrence are displayed next to the alarm list items.
Alarm Recorded parameters
Asystole
Tachy/Brady
Art High/Low
V Fib
V Tachy
V Run >3
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
Alarming level Alarms which start the recording
Severe Asystole, Tachy/Brady, Art High/Low, V Fib, V Tachy