EasyManua.ls Logo

GE Healthcare B30 - Hemodynamic Modules E-PSM, E-PSMP

GE Healthcare B30
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System description
2-31
Synchronization pulse
Pulse width: 10 ms positive pulse
Delay: < 35 ms (R-wave peak to leading edge of pulse)
Amplitude: CMOS compatible
3.5 V min. at 1 mA sourcing
0.5 V max. at 5 mA sinking
Output impedance: 50 ohm
Current limit: 10 mA
Hemodynamic modules E-PSM, E-PSMP
ECG
1
Filter modes:
With 50 Hz power supply frequency:
monitoring filter: 0.5 to 30 Hz
ST filter: 0.05 to 30Hz
diagnostic filter: 0.05 to 150 Hz
With 60 Hz power supply frequency:
monitoring filter: 0.5 to 40 Hz
ST filter: 0.05 to 40 Hz
diagnostic filter 0.05 to 150 Hz
QRS minimum detection level:
Minimum level 0.5 mV with duration between 40 and 120 ms.
Defibrillation protection: 5000 V, 360 J
Recovery time: <5 s
Heart rate:
Measurement range: 30 to 250 bpm
Measurement accuracy: ±5 % or ±5 bpm
Displays average of 10-second median values
2
Display update time: 1 s
Maximum response time of heart rate meter to change in heart rate:
Response time 80 to 120 bpm: 6.9 s
Response time 80 to 40bpm: 8.2 s
Maximum Tall T wave amplitude that does not disturb the heart rate calculation time: >1.4 mV
Input impendance: >2.5 Mohm
The heart rate calculation operates with irregular rhythms of IEC 60601-2-27 6.8.2 bb 4, the heart rate
after a 20 second stablization period as follows:
Figure 101 A1): 80 bpm
Figure 101 A2): 59 bpm
Figure 101 A3): 122 bpm
Figure 101 A4): 117 bpm
Pacemaker pulse detection:
detection level: 2 to 700 mV
pulse duration: 0.5 to 2 ms
Pacemaker detector may not operate correctly during the use of high-frequency (HF) surgical equipment.
The disturbances of HF surgical equipment typically cause false positive pacer detection.
Direct current for leads-off detection
through an active patient electrode: <30 nA
1 The isolation barrier capacitance in the module has been minimized to reduce the hazard of burns in the event of a
defect in the ESU return electrode connection.
2 When the heart rate (HR) changes rapidly, the averaging is restarted.

Table of Contents

Related product manuals