Alarms
4-15
Latching alarms
If the Latching Alarms selection is active, the alarm messages stay on the screen even if the
initial alarm condition goes away. This enables unattended monitoring. You will also hear a
reminder beep every 10 seconds.
To clear the message field of the no-longer active alarm messages and to clear the beep, press
the
Silence Alarms key once.
To clear the message field of all the alarms, enabling only new upcoming alarm messages,
press the
Silence Alarms key twice.
To select latching alarms:
1. Press the
Monitor Setup key.
2. Select Install/Service and enter the password.
3. Select Installation - Alarm Options.
4. Select Latching Alarms - Yes.
Monitor