Cleaning and care
9-5
Modules
Wipe the front panel as any monitor casing.
Do not wash or immerse module in any liquid.
The internal sampling system of the airway module does no cleaning nor sterilization. The
water trap functions as a bacteria filter and there is no reverse flow back to the patient. If the
measuring chamber is suspected to be contaminated (for example, gas zero error), the airway
module should be serviced by authorized service personnel.
Display
Wipe all splashes immediately with a dry cloth.
Wipe the LCD display after use with a cloth moisturized with mild detergent solution.
ECG cables
Wipe the cables with mild detergent solution.
Avoid excessive use of liquids.
Disinfect when necessary.
Allow the product to dry completely after cleaning.
Pulse oximetry sensors
The GE Healthcare pulse oximetry sensors are latex-free. Take possible patient allergies into
account also when selecting the cleaning agent.
1. Detach the sensor from the patient and the monitor.
2. Wipe the sensor with mild detergent solution. Allow it to dry completely before use.
Sensors can be disinfected with chlorite compounds.
The sensors may be sterilized using an ethylene oxide mixture at 50 to 60°C / 120 to 140°F.
NOTE: After ethylene oxide sterilization, sensors must be well aerated in a ventilated place.
WARNING A damaged sensor or a sensor soaked in liquid may cause burns during
electrosurgery.
WARNING To prevent erroneous readings, do not use physically damaged sensors or
sensor cables. Discard a damaged sensor or sensor cable immediately.
Never repair a damaged sensor or cable; never use a sensor or cable
repaired by others.
WARNING Allow the sensor and cable to dry completely after cleaning. Moisture and
dirt on the connector can affect the measurement accuracy.
Temperature probes
Clean with mild detergent solution and rinse with water.
Disinfect or sterilize when necessary.
NIBP cuff hose
The cables and hose can be wiped with mild detergent solution.
Disinfect when necessary.
NIBP cuff
The NIBP cuffs listed in the “Supplies and Accessories” catalog are latex-free. Take possible
patient allergies into account also when selecting the cleaning agent.