Troubleshooting
10-3
Check NIBP NIBP measurement affected by low
blood pressure and pulsation, or a
change in patient's condition.
Check the patient status.
Check the measurement setup.
Check the cuff.
Check sample gas out Sample gas outlet is blocked.
Remove blockage from the sample
gas outlet.
Check SpO2 probe SpO
2
: There is no detectable SpO
2
signal, the sensor is faulty or it is
detached from the patient.
Check the sensor and connections.
Cuff loose Cuff is not attached to the patient or it
is too loose. The hose is not connected
to the module.
Check cuff and hose.
Cuff occlusion Tubes or hose are kinked.
Check tubes and hose.
Cuff overpressure Cuff is squeezed during measurement
and pressure safety limits are
exceeded.
Check cuff, hose and tubes.
Restart measurement.
measurement off Leads have been disconnected for 15
minutes for impedance respiration
check or have just been connected
(<15 seconds ago).
Reconnect the leads or wait to
continue the measurement.
EEPROM error or EPROM error Faulty EEPROM circuit on the CPU
board, or failure in CPU software
memory.
Call service to replace CPU.
Faulty ECG cable The ECG cable is defective.
The ECG cable may be electrically
damaged or the connectors are wet.
Change the ECG cable and the lead
set.
Change the ECG module if the
module connector is wet. Keep the
connectors dry.
Avoid excessive use of liquids when
cleaning cables and connectors.
Gas measurements removed
Reconnect the N-FREC or N-FCREC
module if you want to restart the
CO
2
measurement.
Identical modules There are two identical modules
measuring the same parameter.
You are trying to use two or more
E-PSM(P)W modules or two or more
N-Fx modules at the same time.
You can only use one E-PSM(P)W
and/or one N-Fx module at a time.
Remove extra modules.
Infl. limits! Check setup Adult or child cuff is used but infant
inflation mode is selected.
Check cuff and inflation limits.
InvBP’s not zeroed One or both InvBP channels have not
been zeroed.
Zero the channel indicated or zero
both channels.