EasyManua.ls Logo

GE Healthcare B30 - Page 59

GE Healthcare B30
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System description
2-33
Temperature
1
Measurement range: 10 to 45 °C (50 to 113 °F)
Measurement accuracy: ±0.1 °C (25 to 45 °C)
Probe type: Use only GE Healthcare temperature probes or defibrillator-proof
YSI 400 series probes.
Temperature self-check: At start-up and then every 10 minutes.
Time constant of temperature probes:
Reusable skin temperature probe: 3 s
Reusable adult central temperature probe: 6 s
Reusable pediatric central temperature probe: 4 s
Disposable skin temperature probe: 3 to 6 s
Disposable central temperature probe, 12F: 5 to 8 s
Disposable central temperature probe, 9F: 5 to 8 s
NIBP
Measurement range: adult 25 to 260 mmHg (3.3 to 34.7 kPa)
child 25 to 190 mmHg (3.3 to 25.3 kPa)
infant 15 to 140 mmHg (2.0 to 18.7 kPa)
Typical measuring time: adult less than 30 s, infant less than 25 s
Overall system accuracy: Meets or exceeds SP10-2002 AAMI standards
2
The ESU does not cause a burn hazard through the NIBP cuff, because there is no electrical connection
between the cuff and the NIBP measuring electronics.
NOTE: NIBP measurement is intended for patients who weigh 5 kg (11 lb) and up.
NOTE: The cuff pressure measurement range is equal to cuff nominal and cuff
indication ranges.
Pulse oximetry
SpO
2
1
Automatic scaling of plethysmographic waveform.
Measurement and display range: 0 to 100 %
Calibration range: 70 to 100 %
Calibrated against functional oxygen saturation.
Measurement accuracy
3
: 100 to 70 %, ±2 digits
69 to 0 %, unspecified
Display resolution: 1 digit (1% of SpO
2
)
Wavelength of SpO
2
probe LEDs:
Infrared LED 940 nm
Red LED 660 nm
Maximum energy of SpO
2
probe LEDs:
Infrared LED 42 µJ/pulse
Red LED 62 µJ/pulse
Pulse rate:
Measurement and display range: 30 to 250 bpm
Measurement accuracy
4
: ±5 % or ±5 bpm, whichever is greater
1 The isolation barrier capacitance in the module has been minimized to reduce the hazard of burns in the event of a
defect in the ESU return electrode connection.
2 According to SP10-2002 AAMI 4.4.5.2.B, Intra-arterial method as the reference standard, mean difference of the test
system and the comparison system shall be ± 5 mmHg or less with standard deviation of 8 mmHg or less
3 Accuracy is based on deep hypoxia studies with volunteered subjects during motion and non-motion conditions over a
wide range of arterial blood oxygen saturations as compared to arterial blood CO-Oximetry. Accuracy may depend on
the sensor used, please refer to the instructions for use in the accessory package. The accuracy is expressed as rms. This
means that approximately two-thirds of the data will fall within the accuracy range.

Table of Contents

Related product manuals