B30 Patient Monitor
4-6
Alarm Volume Adjusts the volume of the audio alarms. The range is from 1 (soft) to 10 (loud).
NOTE: Audible alarms cannot be totally silenced with the Alarm Volume function.
Alarm Light
Adjusts the brightness of the alarm light if the light has been enabled (default) through
Monitor Setup - Install/Service - Installation - Alarm Options - Alarm Light- YES. The range
is from 5 to 100%. The light is on during the adjustment to help determine a suitable brightness.
Audio ON/OFF Opens a menu to select Silence Apnea, Silence ECG, Silence Apn & ECG or Silence All.
NOTE: The silencing selections that is, all other selections than the Activate Alarms) are
available only if alarm silencing has been enabled by selecting
Monitor Setup - Install/
Service - Installation - Alarm Options - Show Audio ON/OFF - Yes. By default, it is disabled. For
more details, see ”Enabling or disabling alarm silencing” page 4-14.
Activate Alarms = Activates silenced alarms. This selection is always available for activating
alarms that have been permanently silenced using the Central.
Silence Apn = Silences apnea and disconnection alarms as well as CO
2
, respiration rate.
Silence ECG = Silences arrhythmia alarms and also HR limit alarms.
Silence Apn & ECG = Silences both of above.
Silence ALL = Silences permanently all alarms except FiO
2
<18%, Ppeak >70 cmH
2
O.
Remove Menu Clears the menu selections from the display so that only 10 minute trends and limits are
displayed (push the ComWheel to return the selections on the screen).