Diagnostics/Troubleshooting
7-36 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Hardware
problems
A hardware is an error/malfunction in either the transducer or
the main unit.
A hardware issue may be an error/malfunction on a card.
Software problems Check if a newer software version is available. A software
update may include noise fixes. If needed, update the software.
The following messages may appear during the scan. Please
refer to the cause/action information provided when trying to
resolve the issue.
Table 7-5: Invenia ABUS 2.0 Messages
System Message Cause/Action to Take
[Patient Name] is already selected for this scan You have already selected a patient for this scan.
Abort fininshed. Ready for next scan. You can start the next scan; the previous scan was
aborted.
Aborted The scan has been aborted.
Aborting The current scan is being ended.
Are you sure you want to logout? Prompts the user to confirm that they intended to
stop using the system.
Dataset is not available The selected patient’s dataset is no longer available.
It cannot be re-sent to the workstation.
Imaging Engine Initializing.... The system is in the process of starting up.
Please check the scantype The user must select a scan type before scanning
can begin.
Please input patient information. Prompts the user to fill in all patient information fields
prior to starting the scan.
Please mark all nipple positions The nipple positions on the scan must be set before
proceeding with further action on the system.
Please select a new patient You need to select a new patient in order to begin
the scan.
Please unlock Scan Head Assembly first Login to begin scanning.
Please wait until processing is finished The system is still processing the current task;
please wait to start the next tast.
Press compression button Apply compression before starting the scan.
Press Compression button or Start button Apply compression, then press Start to begin the
scan.
Press Scan Button on Scan Head to Start Scanning Press Start on the Scan Head Assembly to begin the
scan.
Press Start button to start scan Press Start on the Scan Head Assembly to begin the
scan.