Cleaning
Invenia ABUS 2.0 – System Setup and Basic Service Manual 10-7
4700-0043-00 Rev. 4
3. Remove and clean the plastic Clear
Cover of the Scan Head Assembly.
CAUTION: Be sure to properly seat the plastic Clear Cover securely after cleaning.
4. Use disinfectant wipe to clean all portions of the transducer and Scan Head Assembly that comes in
contact with the patient’s breast. See note on approved disinfectant wipes on page 10-5.
Inspect transducer for wear.
CAUTION: DO NOT use an alcohol-based solution to clean the transducer.
Table 10-3: Cleaning the Scan Head (Continued)
Steps Illustrations