Facility needs
Invenia ABUS 2.0 – System Setup and Basic Service Manual 2-9
4700-0043-00 Rev. 4
Important Notices
General Electric, Affiliate, or Distributor Field Engineers and
Application Specialists will setup the system.
NOTICE This medical equipment is approved, in terms of the prevention
of radio wave interference, to be used in hospitals, clinics and
other institutions which are environmentally qualified. The use of
this equipment in an inappropriate environment may cause
some electronic interference to radios and televisions around
the equipment.
Ensure that the following is provided for the new system:
• A separate power outlet with a minimum of a minimum 10
amp circuit breaker for 100-240 VAC.
• Take precautions to ensure that the console is protected
from electromagnetic interference.
Precautions include:
• Operate the console at least 5 meters (15 feet) away
from motors, typewriters, elevators, and other sources
of strong electromagnetic radiation (non-medical grade
UPS must be at least 2 meters (6 feet) away from
console).
• Operation in an enclosed area (wood, plaster or
concrete walls, floors and ceilings) helps prevent
electromagnetic interference.
• Special shielding may be required if the console is to be
operated in the vicinity of radio broadcast equipment.