Site Preparations
2-10 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Important Notices (continued)
The Invenia ABUS 2.0 requires a dedicated power and ground
for the proper operation of its Ultrasound equipment. A separate
power outlet with a minimum 10 amp circuit breaker for 100-240
VAC.
To avoid risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
To avoid risk of fire, the system power must be supplied from a
separate, properly rated outlet.
Under no circumstances should the AC power plug be altered,
changed, or adapted to a configuration rated less than
specified. Never use an extension cord or adapter plug.
To help assure grounding reliability, connect to a “hospital
grade” or “hospital only” grounded power outlet.
Use caution to ensure that the power cable does not
disconnect during system use.
If the system is accidentally unplugged, data may be lost.
To avoid leakage current above safety limits as prescribed by
IEC 60601-1 and to ensure continuity of protective earth. DO
NOT connect Invenia ABUS 2.0 and mains-operated
accessories to a single or multiple socket extension cord or
power strip.